• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD Back to Search Results
Model Number 7060-90
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Arthralgia (2355)
Date of Event 11/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is implant loosening.
 
Event or Problem Description
The patient had right side si joint arthrodesis in (b)(6) 2016 where two implants were placed.The patient did well for several months following the initial procedure but later complained of a recurrence of back pain symptoms.The surgeon determined that the cranial positioned implant may have been loose.In (b)(6) 2018, the surgeon performed a revision surgery where he removed the cranial positioned implant using chisels and replaced it with a wider implant of the same type.The surgeon did not adjust or remove the second preexisting implant.He then added an additional implant between the two existing implants.The status of the patient following the revision procedure is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
MDR Report Key8184843
Report Number3007700286-2018-00110
Device Sequence Number2025026
Product Code OUR
UDI-Device Identifier00859256003323
UDI-Public00859256003323
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/23/2023
Device Model Number7060-90
Device Lot Number493355
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2018
Initial Report FDA Received Date12/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age71 YR
-
-