TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 12/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(6) 2018, in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for erratic blood glucose test results.Daughter is calling on behalf of the customer.The customer is concerned with tests results from back to back blood tests of 408 and 355 mg/dl.Customer was not concerned with high results.The customer's expected fasting blood glucose test result range is 170 - 210 mg/dl.At the time of the call the customer felt well and did not report any symptoms.A past adverse event was reported.On (b)(6) 2018, the customer had not been able to stand and the ambulance had been called.At the hospital the customer was diagnosed with hyperglycemia and was given iv fluids and an insulin drip.The customer's blood glucose test result when at the hospital was 439 mg/dl non-fasting.The customer had been hospitalized from (b)(6) 2018 - (b)(6) 2018.A new medication was suggested; customer was placed on insulin.During the call on (b)(6) 2018, a back-to-back blood test was performed by the customer fasting and produced test results of 310 mg/dl and 323 mg/dl using truemetrix meter.The product is stored according to specification in the living room.The test strip lot manufacturer's expiration date is 02/17/2020, and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8185138
Report Number1000113657-2018-01273
Device Sequence Number16475527
Product Code NBW
UDI-Device Identifier00052569138557
UDI-Public(01)00052569138557
Combination Product (Y/N)N
Initial Reporter StateSC
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/17/2020
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2891
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2018
Initial Report FDA Received Date12/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-