MAUDE Adverse Event Report: DATASCOPE FAIRFIELD SENSATION PLUS 7.5FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE FAIRFIELD SENSATION PLUS 7.5FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0567
Device Problem Failure to Calibrate (2440)
Patient Problem Death (1802)
Event Date 11/24/2018
Event Type  Death  
Manufacturer Narrative
The device was not returned and could not be evaluated.It was discarded by the user.We are unable to confirm the reported event.If new information becomes available, a supplemental report will be submitted.(b)(4).
 
Event Description
It was reported that intra-aortic balloon (iab) therapy was interrupted and an "unable to calibrate sensor" failure was generated.The customer did not changed the iab or iab pump (iabp) and removed the fiber optic cable to allow it to run off of the hemo pressure on semi-auto.Therapy was able to continue successfully.The patient was then transferred to the ccu and an hour later the nurse noted that the machine was alarming.However, the patient expired.The date of death is unknown.The death is not attributed to the device by the facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION PLUS 7.5FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer Contact
15 law drive
fairfield, NJ 07004
MDR Report Key8185770
MDR Text Key131085647
Report Number2248146-2018-00741
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122628
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0684-00-0567
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age YR
Date Manufacturer Received12/03/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age81 YR
Patient Weight67
-
-