| Model Number |
186727980 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that "the patient operated with expedium 5.5 ca 1 year ago.The rod had slipt out of the patient, leaving the construction not satisfying.Patient was fine for the first months after first surgery.Then after a while he got more back problem again.Reoperation shows that some of the sets screw on the distal part of the construction was not that hard to open(seemed like they where not as tight as they should be, final tightening was done by doctors during surgery first time).On the right side the rod was just inside the distal screw, almost on it-s way out as well.Construction was from l2 to s1 with ilium screws bilat.The one ilium screw the whole tulip just had loosen from it screw and was hanging on the rod, without being on the screw itself.During surgery (today) they where tapping with a 10 mm tap before putting in 10x80 viper cortical fix screw in the old ilium fixation (they took out 9x80) the two polyaxial screw driver broke , and the viper screwdriver.The viper part got stuck in the screw while it was half way in the bone, they tried to take i out, but with no luck, leaving it to no other chose than saw of the part that was outside the bone, leaving the rest in the patient.Then they took a exp verse below that (8x80) with success.Final tightening after rod insertion.Some of the sets screws that was taken out had some marks on them".
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Udi: (b)(4).Device evaluated by mfr: corrected.Device manufacture date: additional.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The tulip head and screw shank were returned.It was noted that the shank was cut at its neck.The other half of the shank was not returned.Visual examination found that the second half of the driver¿s tip was lodged inside the drive feature of the screw.The flex ball had fractured.This would allow the saddle and tulip head to separate from the screw shank.The flex ball in the screw may have broken if enough force was applied to it.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.The root cause of the tulip head, saddle, and screw head separating cannot be determined from the sample and the information provided.A potential root cause may be excessive force placed on the polyaxial screw¿s flex ball, especially at an extreme angle in relation to the adjacent components of the screw.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Udi: (b)(4).
|
| |
|
Search Alerts/Recalls
|