| Model Number |
100/897/090CZ |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
Information was received that a smiths medical portex uniperc adjustable flange cuffed tracheostomy tube cuff leaked near the base of the tracheostomy (trach) tube immediately on use.No clinical consequences to the patient were reported.
|
| |
|
Event or Problem Description
|
|
Additional information was received that a tracheostomy (trach) change out was required.
|
| |
|
Additional Manufacturer Narrative
|
|
One tracheostomy was returned evaluation.Visual inspection of the device found it to be in good physical condition.During the manufacturing process the uniperc devices are 100% inflation tested.The inflation test was used as functional testing on the returned sample.After 12 hours, the cuff was noted to be significantly deflated.The sample was then submerged under water, and cuff was inflated using a syringe.Air was observed to be leaking from the place where inflation line is connected with the inflation lumen.The most probable cause of issue was determined to be that the flange was pulled above the last marking, puncturing the lumen.It was determined that the reported customer complaint has been confirmed.
|
| |
|