A steris service technician arrived onsite following the reported event to inspect the 5085 surgical table.The technician identified a hydraulic leak and found that the o-ring on one of the lift cylinders had split, causing the reported event to occur.The technician replaced the o-ring, tested the table, and found it to be operating according to specification.The table was returned to service.The surgical table was installed at the customer's location in 2011 and has been in use for approximately 7 years.It is not under steris service agreement.The user facility's biomed department is responsible for all maintenance activities.No additional issues have been reported.
|