| Model Number |
892109205 |
| Medical Device Problem Codes |
Poor Quality Image (1408); Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Rwmic considers this case open.The user facility and manufacturer will be contacted again in an effort to collect missing information.
|
| |
|
Event or Problem Description
|
|
On (b)(4) 2018, the user facility reported the following to richard wolf medical instruments corporation (rwmic): when scope was inserted into the patient it was cloudy and it had terrible haziness.They had to stop the procedure to switch the scopes.Will the device be returned? yes.Was the device being used during a procedure when the issue occurred? yes.Specifically, was the device being used on a patient when the issue occurred? yes.Was there any injury or illness to patient or other personnel due to issue? no.Did the issue cause a delay in the procedure being performed that put the patient at risk? yes.Was there a similar back-up device available for use? yes.Was the scheduled procedure completed? yes.How was the patient anesthetized? unknown.
|
| |
|
Event or Problem Description
|
|
Follow up #1 to provide investigation results to fda.User facility returned the device to rwmic on december 05, 2018, and was sent to the manufacturer for evaluation.The evaluation was completed on january 25, 2019.The device appeared to be used - there was moisture in the optical system and no leakage on the optic could be found.The reported condition was verified using visual and functional testing.There were no signs of false handling by the user.As no leakage on the optic could be found, it was unclear how the moisture could penetrate into the optical system.The instrument did not meet specifications and a manufacturing defect was not identified.The instrument was repaired.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up #1 to provide fda with device evaluation results.Device labeling was reviewed for patient code and device codes: patient code: not applicable, no patient problem was reported.Device code: ifu was reviewed, before use check irrigation flow, leak tightness and patency (free passage) of the entire system.Caution! increased risk potential if the image is blurred! danger of injuring the patient.Stop the intervention for safety reasons if the image is blurred.Check the image quality of the endoscope before use.Check the glass surfaces for any deposits.Deposits on the glass surfaces can cause a spotted or blurred field of view and thus impair light transmission considerably.Rwmic considers this case open.The user facility and manufacturer will be contacted again in an effort to collect missing information.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up report #2 is to provide fda with new and changed information.Richard wolf medical instruments corporation (rwmic) considers this case closed.In the event that rwmic receives additional information a follow up report will be submitted to the fda.
|
| |
|
Event or Problem Description
|
|
Follow-up report #2 is to provide fda with new and changed information.
|
| |
|