MICROVENTION, INC. AZUR 18 PERIPHERAL V-TRAK SYSTEM; EMBOLIZATION COIL Back to Search Results
Model Number 45-480620
Medical Device Problem Codes Difficult to Remove (1528); Separation Failure (2547); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device was not available to send to the manufacturer for analysis; therefore, a product analysis could not be performed.The root cause is unknown.The instructions for use (ifu) identifies premature or difficult coil detachment as potential complications associated with use of the device.
 
Event or Problem Description
It was reported that after placement of multiple embolization coils in a splenic aneurysm, the last coil came back with the catheter system during withdrawal.Contralateral access was obtained and the coil was retrieved with a snare.A post-procedure ct performed later that evening demonstrated a segment of the coil positioned over the aortic bifurcation with an end in each of the common iliac arteries.The following day, the patient returned for an additional procedure, in which the coil segment was successfully removed with a snare device.There was no patient injury as a result of the issue.The patient is reported to have made a full recovery and is fine now.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZUR 18 PERIPHERAL V-TRAK SYSTEM
Common Device Name
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key8187162
Report Number2032493-2018-00273
Device Sequence Number16518591
Product Code HCG
UDI-Device Identifier00810170017957
UDI-Public(01)00810170017957(11)170817(17)220731(10)170817556
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K090168
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number45-480620
Device Catalogue Number45-480620
Device Lot Number170817556
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/20/2018
Initial Report FDA Received Date12/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-