ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL LINEAR STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number TX30B
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Batch # unk.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that during an unknown procedure, ¿item firing but not sealing tissue.¿ there were no patient consequences reported.
 
Additional Manufacturer Narrative
(b)(4).Batch # p5814y.Device evaluation summary: the analysis results showed that the tx30b device arrived with no apparent damage and with a reload present.The reload was received fully fired.The device was tested for functionality with a test reload and it fired and formed all the staples as intended.The device fired without any difficulties, the staple line was complete, and the staples meet the staple form release criteria.The event described could not be confirmed as the device performed without any difficulties noted.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8187992
Report Number3005075853-2018-15359
Device Sequence Number16804295
Product Code GDW
UDI-Device Identifier10705036003311
UDI-Public10705036003311
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/21/2022
Device Catalogue NumberTX30B
Device Lot NumberP4T21A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2019
Initial Date Received by Manufacturer 11/30/2018
Supplement Date Received by Manufacturer02/26/2019
Initial Report FDA Received Date12/20/2018
Supplement Report FDA Received Date02/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-