MAUDE Adverse Event Report: RADIOMETER MEDICAL APS ABL800 FLEX ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RADIOMETER MEDICAL APS ABL800 FLEX ANALYZER Back to Search Results
Model Number 393-801
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since the unstable calibrations is an on-off issue this indicates that the unstable calibrations are due to a problem with the analyzer and not the membrane unit itself.
 
Event or Problem Description
According to the complaint, a customer obtained the following creatinine results applying the abl800 flex analyzer: (b)(6) 2018/ 11:47h / analyzer serial number (b)(4)/ 0.91 mg/dl / membrane lot r824.At 12:40h at 2018-12-12, the customer changed creatinine membranes, which was the first time since 2018-11-29 at 14:51h.Subsequently, the venous creatinine result was: (b)(6) 2018/ 12:40h/ analyzer serial number (b)(4)/ 4.28 mg/dl /membrane lot r832.Based on the series of measurements, the customer reports the result of 0.91 mg/dl as false low.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABL800 FLEX ANALYZER
Common Device Name
ABL800 FLEX ANALYZER
Manufacturer (Section D)
RADIOMETER MEDICAL APS
aakandevej 21
broenshoej, 2700
DA  2700
MDR Report Key8188732
Report Number3002807968-2018-00068
Device Sequence Number13928405
Product Code CHL
UDI-Device Identifier05700693938011
UDI-Public(01)05700693938011(10)754754R2063N005
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K041874
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number393-801
Device Catalogue Number393-801
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2018
Initial Date Received by Manufacturer 12/14/2018
Supplement Date Received by Manufacturer02/18/2019
Initial Report FDA Received Date12/21/2018
Supplement Report FDA Received Date03/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No