• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO PHILIPPINES CORPORATION MCKESSON MEDI-PAK PERFORMANCE SYRINGE WITH SAFETY NEEDLE; SYRINGE, ANTISTICK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO PHILIPPINES CORPORATION MCKESSON MEDI-PAK PERFORMANCE SYRINGE WITH SAFETY NEEDLE; SYRINGE, ANTISTICK Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient involvement.Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device has not been returned to the manufacturing facility for evaluation.Therefore, the investigation was based upon evaluation of user facility information, and evaluation of retention sample.Based on the results of our investigation, the cause of the complaint could not be identified since condition of the actual sample was not confirmed.Retention samples were visually in good condition and passed the adhesive hold test.We have adhesive hold test to check bonding strength between needle and hub during needle assembly.In addition, the epoxy condition of the needles is visually inspected.Thus, condition of product is assured prior supply to sg production.We have in-process inspection for visual, sensory test and functional test to assure quality performance of assembled sg2 needle.Prior shipment, qc conducts outgoing inspection to assure lots are in good quality.All samples passed.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of terumo (philippines) corporation (manufacturer) registration no.3003902955.Exemption number e2015017.
 
Event or Problem Description
The user facility reported that the syringe with safety needle safety mechanism is faulty.The safety mechanism did not catch and the needle fell out as soon as it was opened.There was no patient impact it was pretreatment.The staff was peeling open unused product prepping for a flu vaccine.There was no procedure impact as the product was not used for the procedure.Additional information received on (b)(6) 2018: it was reported that the needle (steel from needle hub) fell out as soon as it was opened.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the device returned date and to provide the completed investigation results.Twenty pieces of unused same lot samples were returned to the manufacturing facility for evaluation.No defects were found that may result to the complaint.Based on the results of our investigation, the cause of the complaint could not be identified.Received unused same lot samples were visually inspected and no defect was found that may result in detached cannula.Both retention samples and received unused same lot samples were visually in good condition and passed the adhesive hold test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON MEDI-PAK PERFORMANCE SYRINGE WITH SAFETY NEEDLE
Common Device Name
SYRINGE, ANTISTICK
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna
RP 
MDR Report Key8189015
Report Number3003902955-2018-00051
Device Sequence Number19493199
Product Code MEG
UDI-Device Identifier30612479179067
UDI-Public30612479179067
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K051865
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model NumberN/A
Device Catalogue Number102-SN3C231S
Device Lot Number171020A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2018
Initial Date Received by Manufacturer 12/03/2018
Supplement Date Received by Manufacturer12/27/2018
Initial Report FDA Received Date12/21/2018
Supplement Report FDA Received Date01/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-