• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 SROM*STM ST,30+4L NK,18X13X160; S-ROM HIP SYSTEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 SROM*STM ST,30+4L NK,18X13X160; S-ROM HIP SYSTEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 563518
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Test Result (2695); No Code Available (3191)
Date of Event 03/27/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Initial reporter occupation: non-healthcare professional ¿ attorney.(b)(4).
 
Event or Problem Description
Complaint description: patient was revised to address pain.Litigation alleges the patient suffers from discomfort, weakness, loose feeling or popping, difficulty ambulating, and soft tissue/inflammatory response.After review of the medical records for mdr reportability, the revision operative note indicated pain and malpositioned cup (not revised).Ppf alleges abductor muscle repair and elevated metal ions.After review of medical record for mdr reportability, patient was revised to address pain.The cup was in about 5 degrees of anteversion.The stem was in about 10 degrees of anteversion.Doi: (b)(6) 2012 - dor: (b)(6) 2014 (right hip).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8189244
Report Number1818910-2018-78980
Device Sequence Number19493201
Product Code JDI
UDI-Device Identifier10603295178217
UDI-Public10603295178217
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K851422
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2014
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2015
Device Catalogue Number563518
Device Lot Number3221771
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/29/2018
Supplement Date Received by Manufacturer03/20/2019
Initial Report FDA Received Date12/21/2018
Supplement Report FDA Received Date04/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
-
-