• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number ATG80166
Medical Device Problem Codes Retraction Problem (1536); Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The results of the anticipated device evaluation will be provided upon completion of the event investigation.
 
Event or Problem Description
It was reported that during a venous recanalization procedure in the iliac vein the pta balloon allegedly detached from the catheter shaft during removal from the patient.The detached material could not be snared and a stent was used to pin the detached material to the vessel wall.There was no reported patient injury.
 
Event or Problem Description
It was reported that during a venous recanalization procedure in the iliac vein the pta balloon allegedly detached from the catheter shaft during removal from the patient.The detached material could not be snared and a stent was used to pin the detached material to the vessel wall.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found the balloon to be fully detached and was not returned.Additionally, the inner guidewire lumen was stretched distally, indicating retraction issues.Therefore, the investigation is confirmed for balloon detachment as well as for retraction issues.The investigation is also confirmed for a dislodged marker band.The marker band likely dislodged as the inner lumen stretched and as the balloon detached distally.It is likely that the balloon detachment resulted in the retraction issues through the sheath.However, the definitive root cause for the identified balloon detachment could not be determined based on the available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(2923 - device code + descrption3).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8189338
Report Number2020394-2018-02236
Device Sequence Number11237877
Product Code LIT
UDI-Device Identifier00801741060830
UDI-Public(01)00801741060830
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K122984
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 03/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberATG80166
Device Lot Number93UC0206
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2019
Initial Date Received by Manufacturer 12/03/2018
Supplement Date Received by Manufacturer03/08/2019
Initial Report FDA Received Date12/21/2018
Supplement Report FDA Received Date03/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-