• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number AT120124
Medical Device Problem Codes Break (1069); Retraction Problem (1536)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date 03/2020).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the left subclavian, the pta balloon allegedly detached within the sheath.The sheath and detached material were removed together, as one unit.No further treatment was needed.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation, stuck in the user¿s introducer sheath.A visual inspection found the balloon material bunched at the distal end of the sheath, and the sheath tip was found to be flared, indicating retraction issues.The identified stretching to the outer catheter, along with bunching, further indicated retraction issues.The device was distally removed from the sheath, exposing a complete break in the outer catheter shaft at the proximal balloon joint.Therefore, the investigation is confirmed for sheath-related retraction issues, and an outer catheter shaft break.However, the investigation is unconfirmed for the reported device detachment, as the inner polyimide tubing remained intact.It is likely that the reported retraction issues contributed to the identified catheter break.However, the definitive root cause for the reported retraction issue could not be determined based upon available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the left subclavian, the pta balloon allegedly detached within the sheath.The sheath and detached material were removed together, as one unit.No further treatment was needed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8189653
Report Number2020394-2018-02249
Device Sequence Number13928868
Product Code LIT
UDI-Device Identifier00801741062605
UDI-Public(01)00801741062605
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K120971
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAT120124
Device Catalogue NumberAT120124
Device Lot NumberGFBQ3145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2018
Initial Date Received by Manufacturer 12/05/2018
Supplement Date Received by Manufacturer03/07/2019
Initial Report FDA Received Date12/21/2018
Supplement Report FDA Received Date03/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
Patient Weight66
-
-