| Model Number |
SSNF |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Cardiac Perforation (2513)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device was not available for evaluation.The device history was reviewed and found no anomalies.Proper materials and methods of manufacturing were utilized.There is no indication of a relation between the adverse event and the performance of the device.If additional information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that when the transseptal puncture was performed during an atrial fibrillation ablation procedure utilizing the safesept needle free transseptal guidewire along with a competitor dilator and introducer sheath to gain access to the left atrium, the dilator punctured the left atrial appendage.As a result, heart surgery with a sternotomy was performed to close the atrial appendage.No further patient complications have been reported as a result of this event.
|
| |
|