| Catalog Number |
2N3348 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was no flow through a micro-volume extension set.This event occurred before patient use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Pressure test, clear passage test, and functional tests were performed with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|