| Catalog Number |
0279251101 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device experienced continuous activation.It was also reported that the case was completed successfully with no adverse consequences.
|
| |
|
Event or Problem Description
|
|
It was reported that the device experienced continuous activation.It was also reported that the case was completed successfully with no adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Alleged failure: probe continued to activate under coag despite not being used the failure identified in the investigation is consistent with the complaint record.The probable root cause could be the probe short (possibly due to fluid ingress), inadequate gluing operations (tip and core/lumen) or inadequate gluing/welding of the core to handle.
|
| |
|