| Model Number |
R SERIES |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "poor pad contact" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed during review of the device history logs.However, the reported problem could not be duplicated after testing including bench handling and full pads functional testing with the device.The bridge harness was replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|