AV-TEMECULA-CT ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM Back to Search Results
Catalog Number 1012534-40
Medical Device Problem Codes Premature Activation (1484); Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no indication of a lot specific product issue.The investigation determined the difficulties were likely due to operational context.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified lesion in the superficial femoral artery.As the absolute pro self expanding stent system (sess) was being advanced onto an unspecified guide wire that was at the target lesion, resistance was noted and it felt like the sess was sticking to the guide wire.The physician could not remember if guide wire had been wiped down so the sess was attempted to be removed.There was resistance with the guide wire during removal of the sess.As the sess was being backed off of the guide wire, the sheath pulled back and the distal part of the stent deployed and became flowered.Reportedly, the stent was confirmed to not be exposed prior to loading the sess onto the guide wire.Another absolute pro sess and guide wire were used to successfully complete the procedure.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8191390
Report Number2024168-2018-10008
Device Sequence Number2069800
Product Code NIO
UDI-Device Identifier08717648175985
UDI-Public08717648175985
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number1012534-40
Device Lot Number8051861
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2018
Initial Report FDA Received Date12/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
-
-