Model Number 98430
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A lot sample from the customer has been returned and is being processed for evaluation.The device history record for lot aa7261r01 was reviewed and the product was produced according to product specifications.All information reasonably known as of 21 dec 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported that the sutures provided in the kit snapped during the procedure.No further information provided.
 
Additional Manufacturer Narrative
One unused kit was returned by the customer.The sutures were removed from the kit and were found to be intact, with appropriate knots in place.Each suture was firmly pulled at various points.None of the sutures broke.The reported event could not be reproduced.The complaint is not confirmed, and root cause of the reported event remains unknown.All information reasonably known as of 09 jan 2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 16FR DILATOR
Common Device Name
DH EF PERC PLACEMENT PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key8191479
Report Number9611594-2018-00235
Device Sequence Number2118979
Product Code KGC
UDI-Device Identifier10680651984309
UDI-Public10680651984309
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K080253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2019
Device Model Number98430
Device Catalogue Number991098430
Device Lot NumberAA7261R01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 10/30/2018
Supplement Date Received by Manufacturer12/30/2018
Initial Report FDA Received Date12/21/2018
Supplement Report FDA Received Date01/10/2019
Patient Sequence Number1
-
-