| Catalog Number |
99400-003486 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had displayed a "service" message in its display.This message can be indicative of a device lock-up condition.As a result, defibrillation therapy may not be available, if it is needed.There were no reports of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to physio-control for evaluation.A third-party service agent performed an initial evaluation of the customer's device and verified the reported issue.Additionally, physio-control contacted the customer and confirmed that the date that the reported issue was observed was 12/05/2018.As a result, b3, event date, has been updated accordingly.Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.Device not evaluated by manufacturer.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had displayed a "service" message in its display.This message can be indicative of a device lock-up condition.As a result, defibrillation therapy may not be available, if it is needed.There were no reports of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The third-party service agent replaced the system pcb assembly and subsequently observed proper device operation through functional and performance testing.The device was then returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device had displayed a "service" message in its display.This message can be indicative of a device lock-up condition.As a result, defibrillation therapy may not be available, if it is needed.There were no reports of patient use associated with the reported event.
|
| |
|