• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GUIDE BOLT WRENCH; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GUIDE BOLT WRENCH; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71631140
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2018
Event Type  malfunction  
Event Description
It was reported that during surgery, in the tightening process, the round head broke.There was a delay of between 0 and 30 minutes, a backup from a different company was used.No patient injuries were reported.
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The hex end fractured off the device and was returned.The device was manufactured in 2016 and exhibits signs of significant wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDE BOLT WRENCH
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8191783
MDR Text Key131229305
Report Number1020279-2018-02874
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71631140
Device Lot Number16DM16244
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2018
Initial Date Manufacturer Received 11/27/2018
Initial Date FDA Received12/21/2018
Supplement Dates Manufacturer Received11/27/2018
Supplement Dates FDA Received01/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-