• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SINGLE STERILE XCOATED DELPHIN; CENTRIFUGAL PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SINGLE STERILE XCOATED DELPHIN; CENTRIFUGAL PUMP Back to Search Results
Model Number 164275X
Medical Device Problem Code Noise, Audible (3273)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that during priming (after 1-2 minutes) the user observed grinding noise generated by the sarns centrifugal pump head.No consequences or impact to patient.Product was not changed out.Procedure was completed successfully.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on dec 21, 2018.(b)(4).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.(b)(4).Visual inspection was performed on the sample upon receipt, during which no anomalies were noted anywhere on the device.The sample was built into a saline circuit and then set up on a sarns drive motor.The rpm were set at 900 and ramped up by 300 rpm every ten minutes for one hour for a maximum of 2400 rpm.During each interval, the returned sample was observed for any running sound anomalies.No sound anomalies or abnormalities with the functionality of the pump were observed during the test.A retention sample from the affected product code/lot number was obtained and tested the same way, no sound anomalies noted either; however, the issue experienced by the customer is likely due to an abnormal interaction between the seal rotor and stator surfaces.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE STERILE XCOATED DELPHIN
Common Device Name
CENTRIFUGAL PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key8191785
Report Number1124841-2018-00340
Device Sequence Number13943487
Product Code KFM
UDI-Device Identifier00699753450325
UDI-Public(01)00699753450325
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K112229
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model Number164275X
Device Catalogue NumberN/A
Device Lot NumberVL26
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/12/2018
Supplement Date Received by Manufacturer01/11/2019
02/07/2019
Initial Report FDA Received Date12/21/2018
Supplement Report FDA Received Date01/30/2019
02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-