| Catalog Number |
TGU373710 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
11/29/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional devices implanted and/or related to this event: tgu454510/14820255 and tgu373710/14491031.(b)(4).(b)(6).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016, the patient underwent endovascular repair of a thoracic aortic aneurysm using a conformable gore® tag® thoracic endoprostheses.It was reported that on (b)(6) 2018, the patient presented with hemolysis.A type iii endoleak with component separation was identified.There was excessive overlap between the original implanted conformable gore® tag® thoracic endoprostheses (tgu373710/12562071 and tgu373710/14491031) and therefore the tgu454510/14820255 disconnected from both implanted devices.The physician reintervened and bridged the component disconnection by implanting a conformable gore® tag® thoracic endoprosthesis (tgu454520/18712623).The patient tolerated the procedure.
|
| |
|