Brand name: cook spectrum minocycline/rifampin impregnated double lumen central venous catheter.The physician did not use a scalpel to dilate insertion site.Device evaluation: the reported used device was returned to cook inc.For evaluation.Visual and functional testing was performed.Bio matter was present on the device as well as numerous minor kinks.The proximal end had a notably longer flex length when compared to the distal end.The coils along the wire appeared unraveled causing the wire the elongate.Destructive testing was performed and through this, the safety wire was found to have detached at the weld point.Investigation: a document-based investigation reviewed the following: instructions for use, quality control, manufacturing instructions, complaint history and device history records.The device manufacturing record showed no gaps in production or processing controls.A review of the device history record showed two related nonconformances, but all nonconforming product was scrapped and quality control procedures related to the complaint failure mode were performed on all devices in the lot.A complaint history search revealed that there were no other reported complaints for this lot number.Compiling this information, nonconforming product is not suspected in house or in the field.Conclusion: risk assessment determined that risk controls are in place to mitigate this failure mode and that additional risk reduction activities are not necessary at this time.The appropriate personnel were notified, and we will continue to monitor for similar complaints.Based on the information provided, examination of the returned product, and the results of our investigation, a definite root cause could not be determined.Operational use or technique is a potential cause for the deformation.
|
|
It was reported a cook spectrum minocycline/rifampin impregnated double lumen central venous catheter was placed in the subclavian vein.Insertion of the wire guide and catheter were performed with no issue.Upon removal of the wire guide, the wire became "stuck" in the catheter.The physician manipulated and the wire and used force to successfully withdraw the wire.Observation of the wire after removal revealed it was deformed.Cook inc.Received and evaluated the used device.The device was determined to have elongated; therefore, this event is being reported as a malfunction.As reported, the patient did not experience any adverse effects or required an additional procedure due to this occurrence.The scheduled procedure was completed successfully.
|