Model Number A22205C
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to olympus for evaluation.The exact cause of the reported event could not be confirmed, however, based on similar reported complaints the reported event can occur from mechanical overload by the application of excessive force and or the electrode comes into contact (unintended) with other metal parts, e.G.Surgical instruments while the high-frequency output was activated.If the device is returned at a later date, this report will be supplemented accordingly.
 
Event or Problem Description
Olympus was informed that during a transurethral resection of a prostate (turp) procedure, the loop wire at the distal end of the device broke off and fell into the patient¿s bladder.The device was retrieved from the patient.There was no delay in the procedure as the intended procedure was completed using a second device from the same lot.In addition, there was no sparking or arcing observed during procedure; there was no unexpected bleeding to the patient; there was no issue withdrawing the device from the patient and the patient did not require additional procedures or treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE
Common Device Name
HF ELECTRODE, LOOP
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
MDR Report Key8192390
Report Number2951238-2018-00802
Device Sequence Number1616878
Product Code GCP
UDI-Device Identifier14042761036709
UDI-Public14042761036709
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberA22205C
Device Lot Number1000024185
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2018
Initial Report FDA Received Date12/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-