| Model Number |
Z9002 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, explanted, give date: not applicable, as the lens was inserted and removed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the haptic of a z9002 23.0 diopter intraocular lens (iol) was bent when exiting the cartridge.Reportedly, the lens was not stuck in the cartridge.Doctor decided to use another lens, the same size and model.No surgical intervention required.No harm to patient and the patient is fine.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product was not returned to the manufacturing site; therefore, the complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Historical data analysis: a search of complaints revealed no additional investigation requests have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|