Catalog Number 71440822
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/28/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the cutting block ties broke off during the case.No harm to patient or staff reported.There was no backup available, the procedure was completed with no delay.
 
Additional Manufacturer Narrative
The associated complaint device was returned and evaluated.A visual inspection of the returned gii post ref block indicated that the cutting block exhibited wear in each of the cutting slots and damage to the impactor connection point which are consistent with product use.The cutting block bone spikes fractured at the cutting block interface.The fracture of the bone spikes is likely due to stresses in excess of its material properties being applied to the bone spike during use.This device was manufactured in 2013, showing sign of wear/use.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII POST REF A/P BLK SZ 7
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8192744
Report Number1020279-2018-02882
Device Sequence Number2026624
Product Code HRY
UDI-Device Identifier03596010407368
UDI-Public03596010407368
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71440822
Device Lot Number13MM02214
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Initial Date Received by Manufacturer 11/28/2018
Supplement Date Received by Manufacturer11/28/2018
Initial Report FDA Received Date12/21/2018
Supplement Report FDA Received Date01/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-