MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZLB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided.If implanted, give date: unknown, information not provided.If explanted, give date: unknown, information not provided.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zlb00 20.5 diopter intraocular lens (iol) was implanted in the patient's operative eye on an unknown date of service.It was later explanted on an unknown date because the patient's capsular bag ended up not having enough support.The replacement lens was a zma00 20.0 diopter iol.There was no incision enlargement, no vitrectomy, and no sutures used.No additional information was provided.
 
Additional Manufacturer Narrative
The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
 
Additional Manufacturer Narrative
Device evaluation: the product was not returned to the manufacturing site.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.The search revealed that there was another complaint received for this production order number however, it was not related to this complaint.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 12/31/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The lens was inspected with the unaided eye showing it was cut most probably to aid in explant.The lens was cleaned with purified water and dried with compressed air to ease inspection.Further inspection under magnification did not show any damage to the haptics.The complaint event could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8192822
Report Number3011852734-2018-00239
Device Sequence Number16463587
Product Code MFK
UDI-Device Identifier05050474560680
UDI-Public(01)05050474560680(17)210913
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/06/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/13/2021
Device Model NumberZLB00
Device Catalogue NumberZLB00U0205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/31/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/23/2018
Supplement Date Received by Manufacturer01/09/2019
01/17/2019
10/25/2020
Initial Report FDA Received Date12/21/2018
Supplement Report FDA Received Date01/11/2019
01/22/2019
11/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Date Report Sent to FDA01/01/2005
-
-