|
WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5MM CORTEX SCREW SELF-TAPPING 32MM; SCREW,FIXATION,BONE
|
Back to Search Results |
|
| Model Number |
204.832 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
11/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A review of the device history records has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, the patient underwent open reduction internal fixation (orif) of ankle due to an injury.The open distal tibia fracture was plated with one (1) medial distal tibia plate and one (1) locking compression plate (lcp) plate.The fibula was also plated with one (1) one-third (1/3) tubular plate.The patient had an infection due to the open fracture and was made aware on (b)(6) 2018.A hardware removal was conducted on (b)(6) 2018 for all plates including one (1) 3.5mm locking screw self-taping with stardrive(tm) recess 50mm, one (1) 3.5mm locking screw self-taping with stardrive recess 48mm, one (1) locking screw self-taping with stardrive recess 42mm, one (1) 3.5mm locking screw self-taping with stardrive(tm) recess 40mm, one (1) 3.5mm locking screw self-taping with stardrive(tm) recess 18mm, two (2) 2.7mm cortex screw self-tapping 32mm, one (1) 2.7mm cortex screw self-tapping 36mm, two (2) 3.5mm cortex screw self-tapping 38mm, one (1) 3.5mm cortex screw self-tapping 36mm, one (1) 3.5mm cortex screw self-tapping 34mm, five (5) 3.5mm cortex screw self-tapping 32mm, and one (1) 3.5mm cortex screw self-tapping 30mm.Irrigation and debridement was performed during removal surgery.There was no surgical delay.Procedure outcome was successfully completed.Patient outcome is unknown.This report is for one (1) 3.5mm cortex screw self-tapping 32mm.This is report 6 of 10 for (b)(4).(b)(4) captures 10 devices out of 26 and houses 10 impacted products of the 17 unique part and lot number devices.(b)(4) captures 10 devices and (b)(4) captures 6 devices.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device evaluation: investigation flow: adverse event (no reported product problem) visual inspection: the returned device was examined and minor wear/surface abrasions consistent with implantation and explantation were noted; the received condition would not impact the device functionality.No defects or deficiencies were identified.The allegation of adverse event (infection) cannot be confirmed.Conclusion: the received device was found to be without defect or deficiency which would impact functionality.During this investigation no product design or manufacturing issues were observed that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|