• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DRILL, FLEXIBLE, CANNULATED, 8.5MM; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DRILL, FLEXIBLE, CANNULATED, 8.5MM; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number 72203164
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that sa was drilling femoral tunnel.She had just entered the femoral tunnel in forward full speed and was moving the drill slowly.She stopped drill to read the depth off of reamer.The surgeon noticed that the distal tip of reamer between tip and flexible shaft was being stretched.The surgeon commented, "once you start to drill, its going to snap again." and, that's exactly what happened.This is the second flexible reamer to snap in past six weeks.We went through all checkpoints of possible reasons for this and concluded it had to be an issue with the reamer.The reamer has only been used less than 5 times.A kelly was used to remove pieces and a backup device was available to complete the procedure with no significant delay or patient injuries.
 
Additional Manufacturer Narrative
Visual inspection and functional testing could not be performed because the device in question has not been returned for evaluation.Thus, the complaint could not be verified and a root cause could not be determined with confidence.It is difficult to perform an accurate evaluation of a failure without having the failed product.As such the complaint is being closed without conclusion.However, if the product is returned in the future the complaint can be reopened and evaluated.No further investigation is required.
 
Event or Problem Description
It was reported that sa was drilling femoral tunnel.She had just entered the femoral tunnel in forward full speed and was moving the drill slowly.She stopped drill to read the depth off of reamer.The surgeon noticed that the distal tip of reamer between tip and flexible shaft was being stretched.The surgeon commented, "once you start to drill, its going to snap again." and, that's exactly what happened.They went through all checkpoints of possible reasons for this and concluded it had to be an issue with the reamer.No patient injuries were reported but it is unknown if there was a delay or a backup device available to complete the procedure.Attempts were made to retrieve further information but no response was received from the complainant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL, FLEXIBLE, CANNULATED, 8.5MM
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8193056
Report Number1219602-2018-01770
Device Sequence Number268462
Product Code NBH
UDI-Device Identifier03596010654571
UDI-Public03596010654571
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72203164
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/11/2018
Supplement Date Received by Manufacturer02/08/2019
Initial Report FDA Received Date12/22/2018
Supplement Report FDA Received Date02/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-