| Catalog Number |
VBC081002 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Perforation of Vessels (2135)
|
| Date of Event |
12/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Imaging evaluation results.Intended use / indications state, the gore® viabahn® endoprosthesis is indicated for improving blood flow in patients with symptomatic peripheral arterial disease in superficial femoral artery lesions with reference vessel diameters ranging from 4.0 ¿ 7.5 mm.The gore® viabahn® endoprosthesis is indicated for improving blood flow in patients with symptomatic peripheral arterial disease in iliac artery lesions with reference vessel diameters ranging from 4.0 ¿ 12 mm.W.L.Gore & associates has insufficient clinical and experimental data upon which to base any conclusions regarding the effectiveness of the gore® viabahn® endoprosthesis in applications where the endoprosthesis may experience repeated and extreme flexion, such as across the popliteal fossa and the antecubital fossa.Clinical conditions such as excessive bending, tortuosity, and / or repeated and extreme flexion may result in compromised performance or failure of the endoprosthesis.
|
| |
|
Event or Problem Description
|
|
It was reported to gore that a 8mm x 10cm gore® viabahn® endoprosthesis was to be implanted for treatment of the splenic artery aneurysm.Because of the excessively tortuous vessels, resistance was reportedly felt during device advancement.The splenic artery aneurysm was ruptured due to the excessive force used during attempts to get the viabahn device to treatment site.The constrained gore® viabahn® endoprosthesis was retracted out of the patient.Coiling was done to stop the bleeding and the patient was transferred to the icu for surgical intervention.The patient tolerated the procedure.
|
| |
|
Search Alerts/Recalls
|