| Model Number |
9583 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Event date: no event date provided so first date of the month of the aware was used: (b)(6) 2018.Device is a combination product.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 2.25x12mm promus premier stent was selected for treatment.However, the physician bent the hypotube to the stent.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 2.25x12mm promus premier stent was selected for treatment.However, the physician bent the hypotube to the stent.No patient complications were reported.Bsc would like to withdraw this event as the report was sent in error and does not meet the reporting criteria guidelines.The complaint is for a shaft kink with no allegation of stent damage.
|
| |
|
Additional Manufacturer Narrative
|
|
Event date: no event date provided so first date of the month of the aware was used - (b)(6) 2018.Device is a combination product.Correction: a correction was made to the "describe event or problem" section.Bsc would like to withdraw this event as the report was sent in error and does not meet the reporting criteria guidelines.The complaint is for a shaft kink with no allegation of stent damage.Product analysis further confirmed there was no sign of damage, stretching or lifting of the stent struts but did verify multiple severe hypotube kinks.
|
| |
|