• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. DOUBLE FIX KNOTLESS PK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. DOUBLE FIX KNOTLESS PK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72290143
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that during articular labrum repair/arcr procedure, the screw part of the inserter broke off.Pieces were retrieved from the patient.There was a 30 minute delay and the procedure was finished with the same device.
 
Additional Manufacturer Narrative
The reported doublefix knotless device, intended for use in treatment, was not returned for evaluation.A relationship between the product and reported incident cannot be established as the product was not returned.Without the reported product a visual evaluation cannot be performed and customer¿s complaint cannot be confirmed.An exact root cause cannot be determined without evaluation of the device; however, factors unrelated to the manufacture or design of the device that could have contributed to the reported event include: (1) misalignment of inserter (2) over tensioning the suture.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
 
Additional Manufacturer Narrative
The initial incident report was submitted with incorrect manufacturing site registration number (b)(4)(s+n mansfield).The correct manufacturing site registration number is (b)(4)(s+n austin).Corrected data: g1 contact information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLE FIX KNOTLESS PK
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key8194394
Report Number1219602-2018-01782
Device Sequence Number18893183
Product Code MBI
UDI-Device Identifier00885556722497
UDI-Public00885556722497
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153669
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72290143
Device Catalogue Number72290143
Device Lot Number2020671
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/13/2018
Supplement Date Received by Manufacturer02/22/2019
02/08/2020
Initial Report FDA Received Date12/24/2018
Supplement Report FDA Received Date02/25/2019
02/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-