Per health professional/ end user/customer, complaint sample was not available for evaluation.Galt performed investigation and evaluation based on documentation, device history record and complaint history record review.The manufacturing documentations indicates that the purchased materials passed galt incoming inspections requirements, and the dhrs, drawing confirmed all parts were manufactured to specifications without any abnormality or deviations indicated.The information below was acquired from random selected samples for subcomponent crw-010-14 and finished good gwa-038-07 found in galt inventory: distal tip flatten end thickness was within specifications of 0.003" per galt design controls.Tensile pull force for ball weld joint at distal tip is higher than minimum required per engineering drawings and iso: 11070-2014.(galt min.1.55 lbs.), and average results was (2.397 lbs.).Average tensile pull force @ 1cm from uncoiled distal tip core wire without ball weld joint was 2.077 lbs.Average tensile pull force @ 4cm from uncoiled distal tip core wire without ball weld joint was 3.665 lbs.Average torquability for random selected coiled tip samples was 18.7 revolutions @360 degree, that required breaking core wire.Per coronary and cerebrovascular guidewire guidance, january 1995.This investigation provide evidence that galt manufactured the 0.014" x 300cm niti / gold w/ 5cm angled tip in accordance to the specification and galt does not have any history of non-conformances related to the process or design of gwa-038-07 which is used in the finished sterile sgw-060-01.Based on history record for last five years, this type of incident might occur due to end user error.End user might not follow instructions per (instructions for use).Guidewires cannot be removed from patient body through the needle.If any obstructions or resistance occurred, it is recommended to remove entire assembly from patient body not only the guidewires.Removing guidewires through the needle might lead to cut tip of coiled wire by cutting edges of the needles or guidewires might unraveled or tangled while inside patient body causing serious injury.Manufacturer recommend to end user to follow instruction for use stated as warning (do not withdraw guidewire through metal needles; guidewire may shear or unravel).
|