Model Number MMT-7008A
Device Problems Break (1069); Fracture (1260)
Patient Problem Foreign Body Reaction (1868)
Event Date 12/20/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported that sensor did not have the needle when it was removed from the body.The customer's blood glucose level was unknown.The customer reported that the needle was fractured within the body.The sensor will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire st.
northridge, CA 91325-1219
8185464805
MDR Report Key8194635
MDR Text Key131400043
Report Number2032227-2018-78597
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeTU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/20/2018
Initial Date FDA Received12/25/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-