• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAP
Medical Device Problem Codes Obstruction of Flow (2423); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Customer reported via phone call that the insulin pump had damage.Customer's blood glucose level was unknown at the time of the incident.The customer was assisted with troubleshooting.Customer stated that the insulin pump had battery life diminished, o ring from reservoir area missing and insulin pump had random no delivery alarms.The device will not be returned for analysis.
 
Additional Manufacturer Narrative
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform rewind test, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and displacement test or verify no delivery alarm due to completely broken reservoir tube lip.Device had cracked battery tube threads, missing reservoir tube o-ring.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8194832
Report Number2032227-2018-78776
Device Sequence Number2016142
Product Code OYC
UDI-Device Identifier00643169202337
UDI-Public(01)00643169202337
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAP
Device Catalogue NumberMMT-523NAP
Device Lot NumberB1523NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2019
Initial Date Received by Manufacturer 12/20/2018
Supplement Date Received by Manufacturer03/05/2020
Initial Report FDA Received Date12/25/2018
Supplement Report FDA Received Date03/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Patient Weight122
-
-