| Model Number |
MMT-523NAP |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump had damage.Customer's blood glucose level was unknown at the time of the incident.The customer was assisted with troubleshooting.Customer stated that the insulin pump had battery life diminished, o ring from reservoir area missing and insulin pump had random no delivery alarms.The device will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform rewind test, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and displacement test or verify no delivery alarm due to completely broken reservoir tube lip.Device had cracked battery tube threads, missing reservoir tube o-ring.
|
| |
|