The marathon micro catheter was returned for analysis.When measured, approximately 25cm of the marathon distal floppy segment was found to be missing.Flash was observed with the marathon micro catheter hub.The missing marathon segment was not returned.The reason for the segment not returning was not provided.The tubing material at the distal broken end exhibited with jagged edges, necking and stretching.Inspection of the remainder of the marathon, apart from the observed separation, there were no other damage or irregularities found with the catheter.The marathon micro catheter was tested with an in-house tapered mandrel.Based on the device analysis and reported information, the customer¿s report of ¿catheter resistance¿ could not be confirmed.The broken end of the marathon exhibited plastic deformation (jagged edges and stretching) which indicated that the catheter separated when exceeding the tensile strength of the tubing material.It is possible high force was used (pushing/pulling); subsequently causing the catheter to separate.If information is provided in the future, a supplemental report will be issued.
|
Medtronic received report of guidewire resistance in catheter.It was reported that the devices were prepared as indicated in the ifu.It was reported that outside of the patient, the mirage guidewire was inserted into the marathon, but was unable to pass.The marathon was replaced and the guidewire was able to be pass normally.The marathon was suspected to have an occlusion.There were no reports of patient injury in association with this event.The marathon was returned for evaluation.During the analysis the marathon was found to be separated.
|