STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0° X3 INSERT 36MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
623-00-36G |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inflammation (1932); Injury (2348)
|
| Date of Event |
11/28/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
An (b)(6)-year-old man, bmi 24, presented to the clinic with complaints of pain and a cracking sensation in his right hip approximately six months, nine years post primary right tha.During the initial procedure, the primary components utilized an accolade #3 stem with a 127-degree neck-shaft angle and +5 36mm l-fit head.Radiographic images of the right hip showed an apparent trunnion fracture with significant asymmetric wear of the polyethylene liner within the acetabular component.Preoperative bloodwork revealed normal leukocytes and normal-range c-reactive protein (crp) and erythrocyte sedimentation rate (esr).Thus, the patient was consented for revision tha for a presumed trunnion fracture.Intraoperative assessment revealed marked heterotopic ossification along the greater trochanter of the right femur, extending into the gluteus maximus, which was excised and debrided thoroughly.Though the femoral head was intact and showed no gross signs of wear, there was complete dissociation from the femoral stem.There were also significant signs of metallosis with metal-stained debris and granulation tissue, which extended deep to the margins surrounding the acetabular shell.Interestingly, the femoral stem demonstrated significant medial trunnion wear, though no true trunnion fracture was noted, contrary to the initial preoperative radiographic assessment.Stem removal was facilitated by an extended trochanteric osteotomy, and definitive reconstruction was performed with a long cementless femoral stem and luque wire fixation of the osteotomy.Intraoperative assessment demonstrated a stable final construct.The patient had an uneventful postoperative recovery and at the time of the 6-week follow-up had completed his outpatient rehab protocol with excellent effect.
|
| |
|
Event or Problem Description
|
|
An 88-year-old man, bmi 24, presented to the clinic with complaints of pain and a cracking sensation in his right hip approximately six months, nine years post primary right tha.During the initial procedure, the primary components utilized an accolade #3 stem with a 127-degree neck-shaft angle and +5 36mm l-fit head.Radiographic images of the right hip showed an apparent trunnion fracture with significant asymmetric wear of the polyethylene liner within the acetabular component.Preoperative bloodwork revealed normal leukocytes and normal-range c-reactive protein (crp) and erythrocyte sedimentation rate (esr).Thus, the patient was consented for revision tha for a presumed trunnion fracture.Intraoperative assessment revealed marked heterotopic ossification along the greater trochanter of the right femur, extending into the gluteus maximus, which was excised and debrided thoroughly.Though the femoral head was intact and showed no gross signs of wear, there was complete dissociation from the femoral stem.There were also significant signs of metallosis with metal-stained debris and granulation tissue, which extended deep to the margins surrounding the acetabular shell.Interestingly, the femoral stem demonstrated significant medial trunnion wear, though no true trunnion fracture was noted, contrary to the initial preoperative radiographic assessment.Stem removal was facilitated by an extended trochanteric osteotomy, and definitive reconstruction was performed with a long cementless femoral stem and luque wire fixation of the osteotomy.Intraoperative assessment demonstrated a stable final construct.The patient had an uneventful postoperative recovery and at the time of the 6-week follow-up had completed his outpatient rehab protocol with excellent effect.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding wear involving a trident liner was reported.The event was confirmed.Method & results: device evaluation and results: could not be performed as no items associated with the event were returned for evaluation.Medical records received and evaluation: a review of the provided medical records by the consulting clinician indicated "complex malposition of the cup in combination with severe heterotopic ossifications have contributed to an overload condition in the arthroplasty bearing with asymmetric poly wear in the liner, secondary trunnion corrosion, gradual loss of taper lock functionality and ultimately femoral head disassociation requiring revision surgery with stem exchange.Does the review identify any procedural related factors that contributed to the event? cup malposition in lack of anteversion and medialized position heterotopic ossifications represent a procedure-related complication of any arthroplasty with frequently patient-related factors contributing as well.Does the review identify any patient related factors that contributed to the event? heterotopic ossifications have patient-related characteristics with frequently genetic predisposition to the problem, as also suggested in this patient.Does the review identify any device related factors that caused or contributed to the adverse event? no device-related factors are associated with any of the implanted devices." device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the reported event was confirmed based on the review of the medical records by the consulting clinician indicating "complex malposition of the cup in combination with severe heterotopic ossifications have contributed to an overload condition in the arthroplasty bearing with asymmetric poly wear in the liner, secondary trunnion corrosion, gradual loss of taper lock functionality and ultimately femoral head disassociation requiring revision surgery with stem exchange."no device-related factors are associated with any of the implanted devices." if additional information becomes available and/or the product is returned, this investigation will be re-opened.
|
| |
|
Search Alerts/Recalls
|
|
|