• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BIOSURE HA 7MM X 30MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BIOSURE HA 7MM X 30MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 72201774
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/16/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the surgery biosure ha screw broke.No patient injury reported.Back up device was available.Broken pieces were removed from the patient's anatomy.
 
Additional Manufacturer Narrative
One 7x30mm biosure ha screw was returned for evaluation.Visual assessment of the screw confirmed the complaint.The screw was returned in four pieces.Approximately 11mm of the distal threads have broken off.The screws major diameter and wall thickness was measured and found to meet print specifications.With the limited clinical details provided regarding patient bone quality, graft type and size as well as tunnel preparation a root cause cannot be determined.This investigation could not identify any evidence of product contribution to the reported complaint.
 
Additional Manufacturer Narrative
Comp.Representative visual inspection and functional testing could not be performed because the device in question has not been returned evaluation.Thus, the complaint could not be verified and a root cause could not be determined with confidence.It is difficult to perform an accurate evaluation of a failure without having the failed product or the pertinent clinical details to consider.As such the complaint is being closed without conclusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSURE HA 7MM X 30MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8196654
Report Number1219602-2018-01787
Device Sequence Number11293892
Product Code HWC
UDI-Device Identifier03596010611703
UDI-Public03596010611703
Combination Product (Y/N)N
PMA/510(K) Number
K080358
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/03/2021
Device Catalogue Number72201774
Device Lot Number50629411
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/16/2018
Supplement Date Received by Manufacturer01/30/2019
02/27/2019
Initial Report FDA Received Date12/26/2018
Supplement Report FDA Received Date01/31/2019
02/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 YR
Patient Weight68
-
-