DEPUY IRELAND - 9616671 ATTUNE REV TIB CEM CONICAL RMR; ATTUNE STEMMABLE TIBIA INSTRUMENTS : REAMERS Back to Search Results
Catalog Number 250620103
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: evaluation of the returned device finds deformation to the reamer.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the attune revision fb tibial drill bit has burrs on it, which has scored the inside of the drill tower.Plunging the drill bit into the drill tower produces small metal shavings.No surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REV TIB CEM CONICAL RMR
Common Device Name
ATTUNE STEMMABLE TIBIA INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key8196904
Report Number1818910-2018-79129
Device Sequence Number19262889
Product Code HTO
UDI-Device Identifier10603295440567
UDI-Public10603295440567
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number250620103
Device Lot NumberNG60812
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/28/2018
Initial Report FDA Received Date12/26/2018
Patient Sequence Number1
-
-