(manufacturer narrative = t, corrected data = f) internal report # (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-55 - user had an inaccurate reference: competitor's meter: the end user is comparing results obtained from trividia's bgm system to the results from a competitor's bgm system.Test strip udi# (b)(4).Note: manufacturer contacted customer on 12/17/2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that they received the replacement product; however, they didn't start to use it yet.No symptoms or medical attention reported since the last call.
|
|
Consumer reported complaint for low blood glucose test results.Daughter is calling on behalf of the customer.The customer is concerned with all the tests results obtained.The expected non- fasting blood glucose test result range is 150 to 200mg/dl.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the bedroom.During the call on 12/03/2018, a blood test was not performed by the customer (declined).The test strip lot manufacturer's expiration date is 03/31/2020 and open vial date is 11/26/2018.The meter memory was reviewed for previous test result history: result 1 :127mg/dl, date: (b)(6) 2020, time:05:26 pm, n/fasting; result 2 :105mg/dl, date: (b)(6) 2020, time:12:56 pm, n/fasting; result 3 :113mg/dl, date: (b)(6) 2020, time:12:54pm, n/fasting; result 4 :106mg/dl, date: (b)(6) 2020, time:12:53pm, n/fasting; result 5 :127mg/dl, date: (b)(6) 2020, time:05:26pm, n/fasting.
|