MEDOS INTERNATIONAL SÃ RL 6.5 HEALIX ADVANCE KNTLS BR; SOFT-TISSUE ANCHOR, BIOABSORBABLE
|
Back to Search Results |
|
| Catalog Number |
222583 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
12/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).Investigation summary: the complaint device was discarded by the customer therefore, device is not available for a physical evaluation.This complaint is not confirmed.No further information regarding the procedure or the device used has been provided to determine a root cause for this failure.A non-conformance search was performed for this part#: 222583, lot#: l918909, combination and no nonconformances were identified.At this point in time, no corrective action is required, and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Event or Problem Description
|
|
It was reported by the sales rep via phone that during a rotator cuff repair procedure the sales rep's 6.5 healix advance knotless broke while threading the anchor outside of the patient.The sales rep stated that the tip of anchor broke off from the rest of the implant.The sales rep stated that the anchor was being used with dynacord and with 2-2.5 sutures with 4-5 tails that were equal in length.The sales rep stated they were using a 5.5/6.5 healix advance awl for this procedure.The case was completed with another like-anchor in the same bone hole with no patient harm, but a one minute delay to open a new anchor.The device was discarded by the customer.The anchor had a piece break off the distal tip during threading.Implantation wasn¿t attempted due to an inability to load suture into the anchor, so bone quality, cleating, anchor axis, stay suture status should all be inconsequential.We couldn¿t thread the anchor, it broken and was discarded without it ever being inside of the patient.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|