Model Number 1365-53-100
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Tissue Damage (2104); No Code Available (3191)
Date of Event 11/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient made an appointment to his doctor after noticing a bump on his thigh.During the surgery, pseudotumors were removed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
After review of medical records patient was revised to addressed bilateral hip mechanical failure of total hip arthroplasty with pseudotumor formation from metal on metal.There was some corrosion between the head and trunnion.This was cleaned using a sponge and irrigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASPHERE M SPEC 12/14 36 +8.5
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key8197700
Report Number1818910-2018-79166
Device Sequence Number13886221
Product Code JDI
UDI-Device Identifier10603295033967
UDI-Public10603295033967
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K082585
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2014
Device Model Number1365-53-100
Device Catalogue Number136553100
Device Lot Number2955015
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/28/2018
Supplement Date Received by Manufacturer12/28/2020
Initial Report FDA Received Date12/26/2018
Supplement Report FDA Received Date01/12/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINNACLE MTL INS NEUT36IDX52OD.; PINNACLE SECTOR II CUP 52MM.; TRI-LOCK BPS SZ 5 HI OFFSET.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient Weight100
-
-