| Model Number |
2477-0000 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the tubing set leaked at the silicone pumping segment during use on a pediatric patient.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer report that the tubing set leaked was confirmed.Visual inspection of the as-received set observed dried blood residue atop the blood filter drip chamber and within one of the two tubing above the blood filter drip chamber.Further inspection of both tubing engagements to their corresponding inlet ports on the top of the blood filter drip chamber identified a small amount of fluid leaking from the engagement of the tubing with the dried blood residue.Slight tugging at the engagement produced separation from the top of the drip chamber component.Additional visual inspection of the separated tubing end observed insufficient areas of expected solvent.The tubing depth was within normal limits, and dimensional analysis of the separated tubing was observed to be within specification.The root cause of the leak and separation during testing was identified as a manufacturing assembly issue of insufficient solvent applied at the engagement of the tubing to the top of the blood filter drip chamber.
|
| |
|
Event or Problem Description
|
|
The customer reported that the tubing set leaked at the silicone pumping segment during use on a pediatric patient.There was no report of patient harm.
|
| |
|
Search Alerts/Recalls
|