• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: exact date unknown/not provided, best estimate between (b)(6) 2017 - (b)(6) 2018.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that patient had refractive surprise/hyperopic miss after the implantation of model zxr00 16.5 diopter iol (intraocular lens) in her left eye (os).The patient continued to complain so an explant was planned.Through follow-up it was learnt the lens was explanted as planned and exchanged for another zxr00 of a higher diopter (18.0).There was no incision enlargement, vitrectomy, or sutures required.The patient was reported to be stable post exchange.No additional information provided.
 
Additional Manufacturer Narrative
The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
 
Additional Manufacturer Narrative
Device evaluation: the device was not returned at the manufacturing site; therefore, product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the device were reviewed and there were no non-conformances with respect to the manufacturing process.The product was manufactured and released according to specification.Historical data analysis: a search in the complaints history revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key8198052
Report Number9614546-2018-01197
Device Sequence Number252609
Product Code POE
UDI-Device Identifier05050474579026
UDI-Public(01)05050474579026(17)211028
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/06/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/28/2021
Device Model NumberZXR00
Device Catalogue NumberZXR00U0165
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/27/2018
Supplement Date Received by Manufacturer01/07/2019
10/25/2020
Initial Report FDA Received Date12/26/2018
Supplement Report FDA Received Date01/15/2019
11/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-