SMITH & NEPHEW, INC. TWINFIX TI 5.0 ULTRABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72200755
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/05/2018
Event Type  malfunction  
Event Description
It was reported that during a shoulder cuff repair surgery the device broke, all pieces were removed.A backup device was available to complete the procedure with no significant delay or patient injuries.
 
Manufacturer Narrative
One twinfix ti 5.0 anchor assembly was returned for evaluation.Visual assessment confirmed the eyelet portion of the anchor is bent and stripped.The proximal threads of the anchor are also damaged.The hex portion of the inserter was dimensionally inspected and found to meet print specification.The condition of the anchor indicates it was subjected to an excessive amount of torque during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX TI 5.0 ULTRABRAID
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key8198168
MDR Text Key131475884
Report Number1219602-2018-01795
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010577986
UDI-Public03596010577986
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K972326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/24/2022
Device Catalogue Number72200755
Device Lot Number50675956
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
Patient Weight35
-
-