| Model Number |
MMT-523LNAB |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported that the insulin pump was damaged.The customer¿s blood glucose level was 149 mg/dl.The customer reported that there was a crack on the lip ring.The customer reported that the reservoir compartment could no longer hold the reservoir.The customer was advised that the insulin pump will need to be replaced.The customer was advised to disconnect from the insulin pump and revert to backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform the displacement test due to completely broken reservoir tube lip.Device had missing reservoir tube o-ring.
|
| |
|
Search Alerts/Recalls
|