| Model Number |
PARADYM RF VR 9250 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
12/03/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2016, the subject implantable cardiac defibrillator (icd) was implanted with a volta 1cr65.On (b)(6) 2018, noise on the right ventricular channel was detected during the follow-up and in the remote follow-up, suggesting some lead issue.The physician decided the replacement of all the system taking into account the patient age and the icd residual longevity.On (b)(6) 2018, the subject icd and volta lead have been explanted and replaced.The physician is assuming that the lead was broken because the patient is practicing gymnastic with weights.The preliminary analysis suggests that the reported event is due to a lead issue.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016, the subject implantable cardiac defibrillator (icd) was implanted with a volta 1cr65.On (b)(6) 2018, noise on the right ventricular channel was detected during the follow-up and in the remote follow-up, suggesting some lead issue.The physician decided the replacement of all the system taking into account the patient age and the icd residual longevity.On (b)(6) 2018, the subject icd and volta lead have been explanted and replaced.The physician is assuming that the lead was broken because the patient is practicing gymnastic with weights.The preliminary analysis suggests that the reported event is due to a lead issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the preliminary analysis of the returned unit, no issue is suspected on the subject icd.
|
| |
|
Event or Problem Description
|
|
On (b)(6)2016, the subject implantable cardiac defibrillator (icd) was implanted with a volta 1cr65.On (b)(6)2018, noise on the right ventricular channel was detected during the follow-up and in the remote follow-up, suggesting some lead issue.The physician decided the replacement of all the system taking into account the patient age and the icd residual longevity.On (b)(6)2018, the subject icd and volta lead have been explanted and replaced.The physician is assuming that the lead was broken because the patient is practicing gymnastic with weights.The preliminary analysis suggests that the reported event is due to a lead issue.
|
| |
|