U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK SPECTRUM; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK SPECTRUM; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number G23763
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Brand name: cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter tray.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that while flushing the catheter included in the cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter tray, discolored fluid was discovered.The catheter was flushed with normal saline prior to insertion, and the user observed brown fluid upon flushing.The catheter did not make patient contact.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Additional information regarding event details has been requested, but is not available at this time.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Correction: the device was not returned.Investigation evaluation.A review of the complaint history, device history record, instructions for use, manufacturing instructions and quality control of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a review of another complaint (pr245946) with a similar device and failure mode was completed.The physical examination in that investigation found that several lumens of the returned devices flushed brownish liquid.Investigators could recreate the reported failure mode.It was concluded that the cause could not be determined.Because of the similarity of the failure mode between the referenced complaint and the current complaint, it is possible that the two events have a similar cause.Additionally, a document-based investigation evaluation was performed.It was concluded that sufficient activities are in place to identify this failure mode prior to device distribution.A review of the device history record found no nonconformances for lot 8993401.A total of three nonconformances were found for complete catheter lot one ic8750194, of which two of these were unrelated to this event.The additional nonconformance involved one device with ¿coating in i.D.¿ and the device was subsequently scrapped.Complete catheter subassembly lot two ic8659845 had three nonconformances, of which two were found to be unrelated to this event.The other nonconformance involved two devices with ¿250, coating in i.D.¿, of which all affected devices were scrapped.Finally, complete catheter lot three ic8659849 recorded two nonconformances that were found to be unrelated to this incident.It should be noted that there was one other complaint associated with lot 8993401, in which the investigation concluded that a cause could not be established.There is no evidence that nonconforming product exists in house or in the field.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: suggested catheter maintenance "the antimicrobial agents minocycline and rifampin contain yell/orange pigments.Some coloration of the spectrum or spectrum glide catheters is normal." instructions for use " prepare the catheter for insertion by flushing each of the lumens.Based on the information provided, no returned product and the results of our investigation, a definitive root cause could not be established.Most likely the device was flushing out excessive abrm coating.The ifu states that, ¿some coloration of the cook spectrum catheters is normal.¿ per the quality engineering risk assessment, no further action is required.The appropriate personnel have been notified and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No new event description information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK SPECTRUM
Common Device Name
FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8199263
Report Number1820334-2018-03892
Device Sequence Number19257180
Product Code FOZ
UDI-Device Identifier00827002237636
UDI-Public(01)00827002237636(17)200323(10)8993401
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K081113
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/23/2020
Device Model NumberG23763
Device Catalogue NumberC-UTLMY-901J-ABRM-CUSTOM-0025
Device Lot Number8993401
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/30/2018
Supplement Date Received by Manufacturer04/05/2019
Initial Report FDA Received Date12/27/2018
Supplement Report FDA Received Date04/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-